Occupancy and Work Within the SAIL Temperature Controlled Room (Room 116B)
📄 Download PDF 📝 Download Word
SAFE WORK PROCEDURE

Use this form to document Safe Work Procedures for hazardous activities and processes. The information in your Safe Work Procedure (SWP) should be supported by a risk assessment.

List the Hazards and Risk Controls as per Risk Assessment

Associated risk assessment reference Hazards Risk controls
RA-A28-SAIL-116B-temperature-controlled-room Carbon dioxide accumulation in a sealed, recirculating air conditioned room - Main door MUST remain open whenever the room is occupied - primary control
- CO2 monitor triggers automatic fresh air intake before levels become unsafe
- Red mushroom panic button installed in the room to sound an alarm
- Evacuate immediately if the CO2 alarm sounds - do NOT investigate first
- If the door must be closed for thermal stability, leave the room
- General SAIL induction and specific 116B induction
RA-A28-SAIL-116B-temperature-controlled-room Failure of the CO2 monitor, alarm or automatic fresh air intake - The door open rule protects you independently of the monitor
- Do NOT occupy the room for extended work with a faulty CO2 system
- Report any fault, sensor error or alarm activation to the supervisor and building manager
- Scheduled function test and calibration by Facilities
RA-A28-SAIL-116B-temperature-controlled-room Working in isolation within a sealed room, particularly after hours - Notify your supervisor or an appropriate person that you are working in the lab
- Confirm the panic button is reachable from your work position before starting
- Carry a mobile phone
- Agree a check-in time for extended or after-hours work
- Do NOT work alone in the room if feeling unwell
RA-A28-SAIL-116B-temperature-controlled-room Door closed or latched while a person is inside - Secure the door open with a hold-open device or wedge before starting work
- Check the room is unoccupied before closing the door
- Verify the door opens from the inside
- Signage on both faces of the door
RA-A28-SAIL-116B-temperature-controlled-room Electric shock from flood water on the laboratory floor, and slips on wet flooring - Check the floor for standing water BEFORE entering
- Do NOT enter or step into standing water - isolate power at the distribution board first
- Socket outlets are RCD protected and mounted above floor level
- Report water ingress or drainage faults immediately

List Resources Required

Including personal protective clothing, chemicals and equipment needed.

Step by Step Instructions for Undertaking the Task

  1. BEFORE STARTING: NEVER work in Room 116B with the door closed. If the door must be closed to hold temperature, you MUST leave the room. NEVER silence, bypass or obstruct the CO2 monitor, fresh air intake or panic button. Evacuate immediately if the CO2 alarm sounds. Do NOT work alone if you feel unwell.
  2. PRE-ENTRY CHECK 1 - Floor: look for standing water or signs of water ingress before entering. If water is present, do NOT enter. Go to the flood water emergency procedure.
  3. PRE-ENTRY CHECK 2 - CO2 monitor: confirm the monitor is powered and not displaying a fault or alarm. If in alarm, do NOT enter and notify the supervisor and building manager. If faulty, blank or showing a sensor error, do NOT occupy for extended work. Report it.
  4. PRE-ENTRY CHECK 3 - Panic button: confirm the red mushroom button is visible, unobstructed and reachable from where you will be working. If equipment blocks it, clear the access before starting.
  5. PRE-ENTRY CHECK 4 - Door: open the main door fully. Fit the hold-open device or wedge so it cannot swing or be pushed closed. Confirm the door can be opened from the inside.
  6. PRE-ENTRY CHECK 5 - Notify: tell your supervisor or another appropriate person that you are working in the lab and state the expected duration. For extended or after-hours work, agree a check-in time. Carry your mobile phone.
  7. PRE-ENTRY CHECK 6 - Equipment: any instrument operated inside 116B has its own SWP. Complete its pre-operation checks as well.
  8. BRIEF EQUIPMENT CHECK (door closed): permitted ONLY for checking, starting or stopping equipment. Confirm the CO2 monitor is not in alarm, enter, perform the check, and leave. Do NOT settle in to work. If the task extends beyond a brief check, open and secure the door, or leave. If you feel any headache, dizziness or drowsiness, leave immediately.
  9. EXTENDED WORK (door open): confirm the door is secured open before starting. This is not optional. Carry out the work per the relevant equipment SWP.
  10. EXTENDED WORK: monitor yourself throughout for CO2 symptoms - headache, dizziness, drowsiness, rapid breathing, difficulty concentrating. If any symptom appears, leave the room immediately, move to fresh air, and report it. Do NOT dismiss it.
  11. EXTENDED WORK: if the CO2 alarm sounds, evacuate immediately. Do NOT allow anyone to close the door while you are inside. Meet any agreed check-in times.
  12. THERMAL STABILITY WORK REQUIRING A CLOSED DOOR: complete all in-room work first with the door open, then LEAVE THE ROOM and close the door from outside. Verify visually that no one else is inside before closing - do NOT assume. Monitor and control the experiment remotely where possible. To return, repeat the pre-entry checks.
  13. COMPLETION: complete or safely pause work per the equipment SWP. Check that no other person remains in the room.
  14. COMPLETION: remove the door hold-open device and close the door. The room is now unoccupied and the door is closed for thermal stability.
  15. COMPLETION: confirm the CO2 monitor is operating and not in fault, and that the panic button remains unobstructed.
  16. COMPLETION: notify your supervisor or check-in contact that you have finished, if a check-in was agreed. Report any fault, alarm activation or water ingress observed during the session.

Emergency Shutdown Procedures

Emergency Procedures for Fires, Spills or Exposure to Hazardous Substances

Clean Up and Waste Disposal Requirements

References Used in the Development of This SWP

E.g. codes of practice.

Competency Required

Qualifications, certificates, licensing, training — e.g. course or instruction.

Staff Approved to Assess Competence for This SWP

Sergio Leon-Saval, Chris Betters, Barnaby Norris

SWP Sign Off Sheet

SWP name and version: Occupancy and Work Within the SAIL Temperature Controlled Room (Room 116B) (Version 2.0)

In signing this section the assessor agrees that the following persons are competent in following this SWP.

Competency sign-off is recorded on the printed copy of this SWP (use the Word/PDF download) or in the laboratory training register.

Name Signature Date Competent Name of Assessor/Authoriser Assessor/Authoriser Signature
         
         
         
         
         

End of Safe Work Procedure